Product Description
Horizon NXT®
Specifications
Mechanism:
Volumetric (non Drop-Counting), Positive Pressure Displacement Reservoir
Flow Rate Range:
0.1 mL/hr to 999.9 mL/hr
0.1 mL/hr to 99.9 mL/hr (Micro Mode)
Volume to be infused:
0.1 to 9999.9 mL
Variable Occlusion Pressure:
75, 100, 200, 300, 400, 500, 750 mmHg (standard mode default is 300 mmHg)
KVO Flow Rate:
3 mL/hr or selected rate (whichever is lower)
Volumetric Accuracy:
±5%
Weight:
11.9 lbs.
Internal Battery:
12 V sealed lead-acid battery
Battery Capacity:
Approximately 3 hours at 125 mL/hr (based on a new, fully charged battery)
Power Requirements:
90-132 VAC, 30 Watt, 47/63 Hz USA (excluding accessories)
180-264 VAC, 30 Watt, 47/63 International (excluding accessories)
Grounding Resistance:
Meets UL Standard 544
Leakage Current:
Meets or Exceeds UL Standard 544
Alarms:
Air-In-Line, Bag Empty, Door Open, Downstream Occlusion, Hold Time Exceeded, Low Battery, Low Flow from Container, Close Roller Clamp, System Error, Upstream Occlusion
Memory:
Permanent data retention after the pump is turned off (unless battery is depleted or disconnected) unless alternative Data Retention Defaults have been selected in the Biomed Options
Tubing:
Horizon® Pump Sets
Features/Modes
- Primary and Piggyback Modes
- Set Time and Volume
- Profile Mode for TPN
- q(x) Mode
- Variable Occlusion Pressure
- Micro Mode
- Epidural Mode