Specifications
Dimensions
Approx. 4.5″W x 2.0″D x 9.2″H
Weight
Approx. 2 lbs.(Including Batteries)
Flow Rate Accuracy
Infusion: linear rate ±3%. Bolus: linear displacement ±3%.
Syringes Options
Monoject® or B-D®; Monoject® or B-D® 20 cc
Flow Rate Range
0-600 mL/hr
Flow Profile
Bolus: Continuous; Infusion: Pulsed Intermittently
Battery
Typical Use: 150 hours; at low battery and typical use: 4 hours
Four “C” Alkaline Cells
Occlusion Force
8 ± 1 lb.
Maximum Occlusion Pressure
11 psi with 60 cc syringe; 20 psi with 20 cc syringe
Occlusion Detection Time
120 seconds at 36 mL/hr (inverse related to flow rate)
Volume Stored on Occlusion
1.1 mL (approx.)
Back Pressure Effect on Accuracy